Keywords:
Artificial Intelligence, MR physics, MR, Physics, Technology assessment, Artifacts, Quality assurance
When applying DRB, the number of averages did not improve image quality. In some cases, it could have the opposite effect. To avoid the signal reduction phenomenon, the number of averages should be smaller than the PAT factor. Instead, increasing phase oversampling to the range of 150%-200% will resolve artifacts. It is also notable that DRB may have an impact on the general texture of the image.