Back to the list
Congress: ECR25
Poster Number: C-16634
Type: Poster: EPOS Radiologist (scientific)
Authorblock: H. Grover, D. Mogel, A. Kagen, N. Rofsky, C. Chen; New York, NY/US
Disclosures:
Hemal Grover: Nothing to disclose
Daniel Mogel: Nothing to disclose
Alexander Kagen: Nothing to disclose
Neil Rofsky: Nothing to disclose
Christine Chen: Nothing to disclose
Keywords: Genital / Reproductive system male, MR, Imaging sequences, Cancer
Conclusion

3T MRI without DLR demonstrated superior SNR and CNR compared to 1.5T MRI with DLR. A stronger negative correlation between SNR and FT was seen with 3T MRI, however the difference was not statistically significant. 

Clinical Relevance/Applications: Although image quality at 1.5T MRI may improve with DLR, it should not replace 3T MRI at this time given the superior SNR and CNR of 3T MRI. 