Results

1.Image Analysis and Comparison(SNR) 

 In Phantom images, SNR of each ROI on T2 axial(Inphase, Water) according to whether or not the self-made carbon fabric fiber tool was 154.42±3.05 and 115.50±8.34, 31.53±1.08 and 21.24±2.27 and showed significant differences (p<0.05). SNR of ROI on T1 axial was 120.93±24.81 and 103.25±8.53 and significant differences(p<0.05)[table]2.

In Chicken leg images, SNR of all ROIs also showed significant differences (p<0.05) [table]3

2. RF Heating Test of Carbon Fabric Fiber Tool

The temperature change measurement of the Carbon Fabric Fiber according to the time change was at least 20.3 °C to 21.1°C, and Mauchly's sphericity test showed that there was no statistically significant difference with p=0.421(p>0.05)[table]4,[fig]10.

3. Qualitative evaluation

Scores of “aliasing artifact” and “noise” were significant differences (p<0.05).There was no statistic difference in and “blur”(p>0.05)[table]5.