Purpose

This work involved comparing paediatric diagnostic reference levels (DRLs) across three UK NHS hospital sites for typical paediatric examinations carried out on a range of diagnostic x-ray systems. Approximately 22,000 patient doses were analysed between 2019 and 2024 (Fig 1).  11,500 (Fig 2) of these were used for comparison against EDRLs.

The aims of this study were

  • To compare Local DRLs to European DRLs where possible.
  • To perform an audit of paediatric doses and establish paediatric DRLs for the most common examinations
  • To perform inter-room comparisons for similar equipment.
  • Identify areas for optimisation.